Title :
Sensitivity of Pulmonary Crackle Parameters to Filter Cut-Off Frequency
Author :
Yeginer, M. ; Kahya, Y.
Author_Institution :
Bogazici Univ., Istanbul
Abstract :
Pulmonary crackles are very important indicators in the diagnosis of pulmonary disorders. Parameters derived from their waveforms, such as initial deflection width and two-cycle duration width, are useful in classifying a crackle and correlating it with various disorders. To obtain a crackle waveform, the background vesicular sound is filtered out and the crackle parameters are very sensitive to filter cut-off frequency. In this study, sensitivity analysis of waveform parameters of simulated crackles superimposed on real vesicular sound is carried out, emphasizing the necessity of precision in determining the filter cut-off frequency.
Keywords :
diseases; lung; medical signal processing; pneumodynamics; filter cut-off frequency; initial deflection width; pulmonary crackle; two-cycle duration width; Cardiac disease; Cardiology; Cardiovascular diseases; Cutoff frequency; Explosives; Filters; Lungs; Pathology; Sensitivity analysis; Stethoscope; Algorithms; Auscultation; Diagnosis, Computer-Assisted; Humans; Lung; Reproducibility of Results; Respiratory Sounds; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Sound Spectrography;
Conference_Titel :
Engineering in Medicine and Biology Society, 2007. EMBS 2007. 29th Annual International Conference of the IEEE
Conference_Location :
Lyon
Print_ISBN :
978-1-4244-0787-3
DOI :
10.1109/IEMBS.2007.4352478