Title :
Design study of a detector insert for high resolution clinical PET imaging
Author :
Tai, Yuan-Chuan ; Laforest, Richard ; Ruangma, Ananya
Volume :
3
fYear :
2003
fDate :
19-25 Oct. 2003
Firstpage :
1714
Abstract :
A new design that takes advantage of an asymmetric geometry was proposed to improve the resolution of PET imaging. The design employs two sets of detectors with different resolutions. The high resolution detectors are placed near the object being imaged and the lower resolution detectors placed far away from the object. This geometry produces a magnified projection of the PET radioisotope distribution onto the detectors far away, allowing one to resolve structures smaller than the intrinsic resolution of the larger detectors. A detector insert based on this new geometry is proposed to allow coincidence between the insert and an existing PET scanner. The device will allow a clinical PET scanner to zoom-in to the regions of interest with improved image resolution. Preliminary data demonstrated significant improvement in image resolution with the new design. Individual components of the proposed device need to be optimized to achieve good sensitivity to support the improved resolution.
Keywords :
gamma-ray detection; image resolution; image sensors; medical image processing; positron emission tomography; PET radioisotope distribution; asymmetric geometry; detector insert; high resolution clinical PET imaging; improved image resolution; Cameras; Detectors; Geometry; Head; High-resolution imaging; Image resolution; Object detection; Positron emission tomography; Sensor arrays; Spatial resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2003 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-8257-9
Type :
conf
DOI :
10.1109/NSSMIC.2003.1352209
Filename :
1352209
Link To Document :
بازگشت