Title :
A Gabor-based technique for bias removal in MR Images
Author :
Ardizzone, E. ; Pirrone, R. ; Mastrella, M. ; Gambino, O.
Author_Institution :
Univ. degli studi di Palermo, Palermo
Abstract :
Magnetic Resonance images are often characterized by irregularly displaced luminance fluctuations that are called bias artifact. This disturb is due to a drop in signal intensity caused by the distance between imaged sample and receiver coil. An original approach to bias removal in Magnetic Resonance images is presented, which is based on the use of Gabor filter to extract the artifact. The proposed technique restores the image using a correction model, which is derived from the attenuation of signal diffusion across the tissues. No hypotheses are made about the structure of the tissues under investigation and the used MR spectrum. The approach is presented in detail, and extensive experimental results are reported along with a comparison with other popular techniques for bias removal.
Keywords :
Gabor filters; biomedical MRI; feature extraction; image restoration; medical image processing; Gabor filter; bias artifact removal; image restoration; luminance fluctuation; magnetic resonance image; receiver coil; signal diffusion; Attenuation; Coils; Fluctuations; Gabor filters; Image restoration; Magnetic fields; Magnetic resonance; Magnetic resonance imaging; Polynomials; Signal restoration; Algorithms; Artifacts; Image Enhancement; Image Interpretation, Computer-Assisted; Magnetic Resonance Imaging; Reproducibility of Results; Sensitivity and Specificity; Signal Processing, Computer-Assisted;
Conference_Titel :
Engineering in Medicine and Biology Society, 2007. EMBS 2007. 29th Annual International Conference of the IEEE
Conference_Location :
Lyon
Print_ISBN :
978-1-4244-0787-3
DOI :
10.1109/IEMBS.2007.4352539