    Position mapping, energy calibration, and flood correction improve the performances of small gamma camera using PSPMT

  • Author

    Jeong, Myung Hwan ; Choi, Yong ; Chung, Yong Hyun ; Song, Tae Yong ; Jung, Jin Ho ; Hong, Key Jo ; Min, Byung Jun ; Choe, Yearn Seong ; Lee, Kyung-Han ; Kim, Byung-Tae

  • Author_Institution
    Dept. of Nucl. Medicine, Sung Kyun Kwan Univ., Seoul, South Korea
  • Volume
    3
  • fYear
    2003
  • fDate
    19-25 Oct. 2003
  • Firstpage
    2103
  • Abstract
    The purpose of this study is to improve the performances of a small gamma camera using position sensitive PMT (PSPMT) by applying position mapping, energy calibration, and flood correction. The small gamma camera consists of a 5" PSPMT coupled with either CsI(Tl) array or NaI(Tl) plate crystals. Flood images were obtained intrinsically in CsI(Tl) array system and with lead hole mask in NaI(Tl) plate system. The position mapping was performed by locating crystal arrays and hole positions in the two systems. The energy calibration was performed using energy discrimination table for each pixel array or for each hole position. The flood correction was performed using a uniformity correction table containing the relative efficiency of each image element. The resolution of the CsI(Tl) array system remained similar before and after corrections. On the other hand, the resolution of the NaI(Tl) plate system was improved about 16% after correction. The uniformity and linearity improved 33.9% to 11.6% and 0.5 mm to 0 mm, respectively, in the CsI(Tl) array system. The corrections more effectively improve the uniformity and linearity in the NaI(Tl) plate system, 21.3% to 9.2% and 0.5 mm to 0 mm, respectively. Furthermore, the resolution deterioration observed in the NaI(Tl) plate system at the outer part of FOV, was considerably diminished after the corrections. The results of this study indicate that the correction algorithms considerably improve the performances of a small gamma camera and the performance gain is more prominent in the system employing a plate type crystal.
  • Keywords
    cameras; energy measurement; gamma-ray apparatus; gamma-ray detection; photomultipliers; position measurement; position sensitive particle detectors; CsI(Tl) array system resolution; FOV; NaI(Tl) plate crystals system resolution; PSPMT; correction algorithms; crystal arrays; energy calibration; energy discrimination table; field-of-view; flood correction; flood images; gamma camera; hole positions; image element relative efficiency; lead hole mask; linearity; performance gain; photomultiplier tubes; pixel array; plate type crystal; position mapping; position sensitive PMT; resolution deterioration; uniformity correction table; Biomedical imaging; Calibration; Cameras; Collimators; Crystals; Energy resolution; Floods; Linearity; Performance gain; Table lookup;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2003 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8257-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2003.1352295
  • Filename
    1352295