  • Abstract
    Large-area, multi-pixel photon counters will be used for the electromagnetic Barrel Calorimeter of the GlueX experiment at Jefferson Lab. These photo sensors are based on a 3 ×3 mm2 cell populated by 50 μm pixels, with 16 such cells tiled in a 4 ×4 arrangement in the array. The 16 cells are summed electronically and the signals are amplified. The photon detection efficiency of a group of first-article units at room temperature under conditions similar to those of the experiment was extracted to be (28 ±2(stat) ±2(syst))%, by employing an analysis methodology based on Poisson statistics carried out on the summed energy signals from the units.
  • Keywords
    Poisson distribution; particle calorimetry; position sensitive particle detectors; GlueX experiment; Jefferson Lab; Poisson statistics; electromagnetic barrel calorimeter; large-area multipixel photon counters; photon detection efficiency; photosensors; size 50 mum; temperature 293 K to 298 K; Arrays; Current measurement; Measurement by laser beam; Noise; Photodiodes; Photonics; Wavelength measurement; Multi-photon pixel counters; photon detection efficiency; silicon photomultipliers;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2015.2442262
  • Filename
    7161418