DocumentCode :
1827730
Title :
An attempt at using TMA to improve the energy resolution of high-pressure Xe ionization chambers
Author :
Bolotnikov, Aleksey ; Ramsey, Brian
Author_Institution :
NASA Marshall Space Flight Center, Huntsville, AL, USA
Volume :
1
fYear :
1996
fDate :
2-9 Nov 1996
Firstpage :
667
Abstract :
Preliminary results of using TMA (trimethylamine) to improve the energy resolution of a high-pressure (>50 atm) Xe ionization chamber are presented. This work borrowed a technique which has been successfully applied to dope liquid Ar and Xe. However, problems related to the contamination of TMA were found to be more difficult to overcome. The maximum density of Xe at which the influence of impurities could be neglected was only 0.3 g/cm3 at a concentration of TMA of about 80 ppm. As in the case of liquid Xe, doping increases the collected charge and considerably improves energy resolution, but only at low fields
Keywords :
gamma-ray detection; ionisation chambers; 50 atm; TMA; collected charge; contamination; energy resolution; gamma detector; high-pressure Xe ionization chambers; impurities; trimethylamine; Cleaning; Coordinate measuring machines; Detectors; Doping; Electric variables measurement; Energy resolution; Gravity; Impurities; Ionization chambers; Q measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
Conference_Location :
Anaheim, CA
ISSN :
1082-3654
Print_ISBN :
0-7803-3534-1
Type :
conf
DOI :
10.1109/NSSMIC.1996.591087
Filename :
591087