Equivalent perception dose of pulsed fluoroscopy
Author :
Aufrichtig, Richard ; Xue, Ping ; Wilson, David L.
Author_Institution :
Dept. of Biomed. Eng. & Radiol., Case Western Reserve Univ., Cleveland, OH, USA
fYear :
1994
fDate :
3-6 Nov 1994
Firstpage :
526
Abstract :
The authors find that pulsed fluoroscopy of stationary objects gives average dose savings of 22%, 38%, and 49% at 15, 10 and 7.5 acq/sec, respectively. Dose savings depend on object size, with less savings for smaller objects. A model that includes the spatio-temporal response of the human visual system describes these results with one free parameter
Keywords :
diagnostic radiography; dosimetry; visual perception; equivalent perception dose; human visual system; medical diagnostic imaging; object size; pulsed fluoroscopy; spatiotemporal response; stationary objects; Filtering; Frequency; Humans; Image sequences; Matched filters; Object detection; Radiology; Signal to noise ratio; Visual system; X-ray imaging;