• DocumentCode
    1828945
  • Title

    Printed sources for positron emission tomography (PET)

  • Author

    Sossi, Vesna ; Buckley, Kenneth R. ; Piccioni, Paul ; Rahmim, Arman ; Camborde, Marie-Laure ; Lapi, Suzy ; Ruth, Thomas J.

  • Author_Institution
    British Columbia Univ., Vancouver, BC, Canada
  • Volume
    4
  • fYear
    2003
  • fDate
    19-25 Oct. 2003
  • Firstpage
    2433
  • Abstract
    We have developed a method that allows manufacturing of 18F radioactive printed sources of any size and shape using a standard ink-jet printer. To test the point sources we used them to measure the resolution of three different tomographs, the Siemens/CTI ECAT 953B, the CPS High resolution tomograph (HRRT) and the Concorde microPET R4. The resolution of these tomographs spans a fairly wide range. Where appropriate, the resolution agrees with published values. A preliminary comparison of the point FWHM obtained with and without additional attenuating material showed no influence of the additional attenuation material on their values, even though the presence of the attenuating material increased the counting statistics by approximately 50%.
  • Keywords
    fluorine; image resolution; positron emission tomography; printing; radioisotopes; 18F radioactive printed sources; CPS High resolution tomograph; Concorde microPET R4; F; HRRT; PET; Siemens/CTI ECAT 953B; attenuating material; counting statistics; point FWHM; point sources; positron emission tomography; printed source size; shape; standard ink-jet printer; Energy resolution; Image resolution; Ink jet printing; Manufacturing; Materials reliability; Positron emission tomography; Radioactive materials; Spatial resolution; Standards development; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2003 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8257-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2003.1352386
  • Filename
    1352386