DocumentCode :
1829370
Title :
A mathematical observer study for evaluation of a model-based compensation method for crosstalk in simultaneous dual isotope SPECT
Author :
Song, X. ; Frey, E.C. ; He, X. ; Segars, W.P. ; Tsui, B.M.W.
Author_Institution :
Johns Hopkins Univ., Baltimore, MD, USA
Volume :
4
fYear :
2003
fDate :
19-25 Oct. 2003
Firstpage :
2521
Abstract :
We have previously developed a model-based method to compensate for the crosstalk in simultaneous acquisition of Tc-99m stress and Tl-201 rest myocardial perfusion SPECT. The purpose of this study is to evaluate the performance of this method in terms of a defect detection task using a mathematical Channelized Hoteling Observer (CHO) study. In this study we also optimized the iteration number in reconstruction of Tl distributions. A population of male and female NCAT phantoms were used to realistically model variation in patients. Both defect-free and defect-present data of Tc and Tl were simulated using the SimSET/PHG codes and our newly developed and validated angular response function (ARF) method to mimic simultaneous acquisition and separate acquisition. Poisson noise was modeled at clinically realistic levels. The Tl data were reconstructed using filtered backprojection (FBP) and ordered subset expectation maximization (OSEM) algorithms for up to 20 iterations with and without the model-based crosstalk compensation (MBC). CHO methodology and receiver operating characteristic (ROC) analysis were applied to short axis Tl images to obtain ROC curves and area under ROC curves (AUC). The AUC values were compared with those from separately acquired data reconstructed using FBP and OSEM algorithms. The results show that a relatively small number of iteration (∼3) is optimal for the postreconstruction filtering cutoff frequency of 0.16 pixel-1 used in this study. Also, the AUC values obtained with the MBC were significantly better than those without crosstalk compensation and those with the FBP reconstruction method applied to the separately acquired data.
Keywords :
compensation; haemorheology; noise; observers; optimisation; phantoms; single photon emission computed tomography; NCAT phantoms; Poisson noise; SimSET/PHG codes; Tc-99m stress; Tl-201 rest myocardial perfusion SPECT; angular response function method; area under ROC curves; crosstalk; defect-free data; defect-present data; female population; filtered backprojection; iteration number optimisation; male population; mathematical Channelized Hoteling Observer; mathematical observer; model-based compensation method; model-based crosstalk compensation; ordered subset expectation maximization algorithms; postreconstruction filtering cutoff frequency; receiver operating characteristic analysis; separate acquisition; short axis Tl images; simultaneous acquisition; simultaneous dual isotope SPECT; Crosstalk; Filtering; Image analysis; Image reconstruction; Imaging phantoms; Isotopes; Mathematical model; Myocardium; Noise level; Stress;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2003 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-8257-9
Type :
conf
DOI :
10.1109/NSSMIC.2003.1352404
Filename :
1352404
Link To Document :
بازگشت