    Feasibility study of using PEImager scanner for positron emission mammography

  • Author

    Jan, Meei-Ling ; Chuang, Keh-Shih ; Ni, Yu-Ching ; Wu, Jay ; Pei, Cheng-Chih ; Yeh, Ching-Kai ; Fu, Ying-Kai

  • Author_Institution
    Inst. of Nucl. Energy Res., Lung-Tan, Taiwan
  • Volume
    4
  • fYear
    2003
  • fDate
    19-25 Oct. 2003
  • Firstpage
    2594
  • Abstract
    The purpose of this work is to study the feasibility of using PEImager scanner for positron emission mammography (PEM). PEM can be performed by using two opposite detectors. The two-detector positron projection imaging has less depth information, because of the limited number of line of responses (LORs). In this work, an iterative back projection algorithm is employed for reconstruction of projection data. Although the number of LORs are limited, the locations and sizes of hot spots in breast phantom still can be determined from the reconstructed images.
  • Keywords
    image reconstruction; iterative methods; mammography; medical image processing; phantoms; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; PEImager scanner; breast phantom; feasibility study; iterative back projection algorithm; line of responses; position sensitive BGO detectors; positron emission mammography; reconstructed images; two-detector positron projection imaging; Breast cancer; Detectors; Image reconstruction; Imaging phantoms; Mammography; Nuclear and plasma sciences; Positron emission tomography; Projection algorithms; Radioactive decay; Whole-body PET;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2003 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8257-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2003.1352421
  • Filename
    1352421