  • Abstract
    This paper presents the design, implementation, and nuclear magnetic resonance (NMR) measurements of a wireless, magnetic compliant, implant that increases the signal sensitivity of NMR images by 3.8dB and 2.6dB in 4.7T and 11.1T magnetic field strengths respectively. The device supports sustained reliable operation through a strongly coupled resonance wireless powering scheme in addition to improving high-resolution image SNR up to 73% within the region of interest.
  • Keywords
    CMOS integrated circuits; artificial organs; biomedical NMR; biomedical electronics; biomedical telemetry; image resolution; medical image processing; patient monitoring; radiotelemetry; CMOS process; NMR compliant wirelessly programmable implant; NMR images; bio-artificial pancreas in vivo monitoring; high-resolution image SNR; magnetic field strengths; magnetic flux density 4.7 T to 11.1 T; nuclear magnetic resonance measurements; region of interest; signal sensitivity; strongly coupled resonance wireless powering scheme; wireless magnetic compliant implant; Coils; Energy measurement; Frequency measurement; Magnetic field measurement; Magnetic resonance imaging; Nuclear magnetic resonance; Signal to noise ratio; Biomedical implant; CMOS; nuclear magnetic resonance; strongly coupled resonance;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    VLSI Circuits Digest of Technical Papers, 2014 Symposium on
  • Conference_Location
    Honolulu, HI
  • Print_ISBN
    978-1-4799-3327-3
  • Type

    conf

  • DOI
    10.1109/VLSIC.2014.6858410
  • Filename
    6858410