A posteriori respiratory motion gating of dynamic PET images
Author
Visvikis, Dimitris ; Barret, Olivier ; Fryer, Tim ; Turzo, Alexandre ; Lamare, Frederic ; Le Rest, Catherine Cheze ; Bizais, Yves
Author_Institution
INSERM, Brest, France
Volume
5
fYear
2003
fDate
19-25 Oct. 2003
Firstpage
3276
Abstract
The presence of patient physiological motion during imaging may cause significant artifacts in image quality. Proposed correction methodologies involve the use of gated acquisitions through simultaneous recording of an external signal. The purpose of our work is to determine the feasibility of post-acquisition synchronization of dynamically acquired PET images in the absence of any external signal. The principle of the technique is based on the assumption that although the amplitude of the motion may vary from pixel to pixel inside the same organ, the frequency of the periodic motion is the same. Under such conditions, the prerequisite for a posteriori gating is the ability to accurately estimate that frequency. We performed simulation studies using the NCAT phantom and a Monte Carlo simulation of the GE Advance PET system (3D mode of operation). A number of NCAT emission and the corresponding transmission frames were generated throughout a respiratory cycle. Time frames of 0.15, 0.45 and 0.62 seconds were simulated with variable count statistics (namely 30k, 70k and 120k of total simulated coincidences). Time activity curves were obtained, for each of the dynamic series formed, using different ROIs and Fourier transform was performed in order to estimate the frequency of the simulated motion. We were able to determine the frequency of motion (within 2% of the simulated frequency) for all three frame time durations evaluated. Using the estimated frequency we were able to calculate on a pixel by pixel basis the amplitude and the phase of the motion, allowing us to reconstruct an a posteriori gated time series.
Keywords
Fourier transforms; Monte Carlo methods; image motion analysis; image reconstruction; medical image processing; phantoms; pneumodynamics; positron emission tomography; time series; 0.15 s; 0.45 s; 0.62 s; Fourier transform; GE Advance PET system; Monte Carlo simulation; NCAT phantom; a posteriori gated time series reconstruction; dynamic PET images; image quality; patient physiological motion; posteriori respiratory motion gating; time activity curves; Amplitude estimation; Fourier transforms; Frequency estimation; Frequency synchronization; Image quality; Imaging phantoms; Motion estimation; Phase estimation; Positron emission tomography; Statistics;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2003 IEEE
ISSN
1082-3654
Print_ISBN
0-7803-8257-9
Type
conf
DOI
10.1109/NSSMIC.2003.1352596
Filename
1352596
Link To Document