• DocumentCode
    1833691
  • Title

    High pressure xenon gamma-spectrometers for “in-situ” geophysical investigation

  • Author

    Dmitrenko, V.V. ; Vlasik, K.F. ; Gratchev, V.M. ; Kondakova, O.N. ; Ulin, S.E. ; Uteshev, Z.M.

  • Author_Institution
    Moscow Eng. Phys. Inst., Russia
  • Volume
    2
  • fYear
    1996
  • fDate
    2-9 Nov 1996
  • Firstpage
    942
  • Abstract
    This paper discusses a design and the main characteristics of a high-pressure xenon cylindrical ionization chamber. It is shown that the energy resolution of the gamma-ray spectrometer is less than 4% at 662 keV, and does not vary widely in the broad temperature range between 20°-175°C. Results are submitted of vibration tests of the cylindrical gamma-ray detector. Discussed are the prospects of high-pressure xenon cylindrical gamma-ray detectors for use in geophysical research
  • Keywords
    gamma-ray detection; gamma-ray spectrometers; geophysical equipment; ionisation chambers; xenon; 20 to 175 C; Xe; gamma-ray spectrometer; geophysical research; high pressure xenon gamma-spectrometers; high-pressure xenon cylindrical ionization chamber; Ceramics; Energy resolution; Gamma ray detectors; Ionization chambers; Isolators; Scintillation counters; Solid scintillation detectors; Voltage; Wire; Xenon;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
  • Conference_Location
    Anaheim, CA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-3534-1
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1996.591503
  • Filename
    591503