Abstract
This review paper presents the current state of the art imaging systems using room-temperature CdTe and CdZnTe X-ray and gamma-ray detectors. Imaging systems composed of both single elements and monolithic segmented CdTe and CdZnTe arrays are used in medical diagnostics, astronomy, research and industry. This paper is divided into three sections: The first section describes different types and configurations of CdTe or CdZnTe detectors used in imaging systems, with the emphasis on single charge carrier collection detectors and front-end read-out electronics in the form of Application Specific Integrated Circuits associated with these detectors. The second section discusses the advantages, disadvantages and limitations with respect to the efficiency and area of the imaging systems using CdTe and CdZnTe detectors compared to other types of X and gamma ray detectors. The third section describes several examples of imaging systems utilizing CdZnTe detectors: 1) A large area imaging telescope for the detection of gamma ray bursts using a coded aperture collimator. 2) A large area medical nuclear camera for cardiology and scintimammography using a parallel hole collimator. 3) A large field of view gamma camera based on a rotating slat collimator and 4) A prototype of an electronically collimated Compton camera.
Keywords
X-ray apparatus; X-ray astronomy; X-ray detection; X-ray imaging; application specific integrated circuits; astronomical telescopes; biomedical equipment; biomedical imaging; cameras; cardiology; collimators; gamma-ray apparatus; gamma-ray detection; mammography; nuclear electronics; readout electronics; reviews; semiconductor counters; Application Specific Integrated Circuits; astronomy; cardiology; coded aperture collimator; electronically collimated Compton camera; front-end read-out electronics; gamma ray burst detection; imaging systems area; imaging systems efficiency; industry; large area imaging telescope; large area medical nuclear camera; large field of view gamma camera; medical diagnostics; monolithic segmented CdTe array imaging systems; monolithic segmented CdZnTe array imaging systems; parallel hole collimator; research; review; room-temperature CdTe X-ray detectors; room-temperature CdTe gamma-ray detectors; room-temperature CdZnTe X-ray detectors; room-temperature CdZnTe gamma-ray detectors; rotating slat collimator; scintimammography; single charge carrier collection detectors; single element imaging systems; Astronomy; Cameras; Charge carriers; Collimators; Gamma ray detection; Gamma ray detectors; Image segmentation; Medical diagnosis; Optical imaging; X-ray imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2003 IEEE
ISSN
1082-3654
Print_ISBN
0-7803-8257-9
Type
conf
DOI
10.1109/NSSMIC.2003.1352603
Filename
1352603
Link To Document