DocumentCode
1837274
Title
Investigation of the new prediction method for Ht values during the plasma exchange
Author
Nitta, Y. ; Akutagawa, M. ; Miyamoto, Hideaki ; Okahisa, T. ; Ohnishi, Y. ; Nishimura, M. ; Nakane, S. ; Kaji, R. ; Kinouchi, Y.
Author_Institution
Univ. of Tokushima 2-1, Tokushima
fYear
2007
fDate
22-26 Aug. 2007
Firstpage
3160
Lastpage
3163
Abstract
There is a disorder that is called Guillain-Barre Symdrome(GBS). Currently, plasmapheresis and high-dose immunoglobulin therapy are used for GBS. Plasmapheresis is one of the blood purifications. When the blood purification is enforced, it should be careful to decrease the blood pressure. In the Tokushima University Hospital, the hematocrit(Ht) values are measured using Crit-Line Monitor(CLM), because the Ht values are rate of red cell in blood, and are relation of the blood pressure. Hence, it is important to predict Ht value using the prediction method. There are various prediction methods, but we proposed the new prediction method. The purpose of this study is the prediction of Ht value after 1,3, and 5 minutes during the plasma exchange using the new prediction method. As the results, rms errors were small in all prediction, but the predicted Ht values are same the measured Ht value before 1,3, and 5 minutes, because the measured Ht values are strong autocorrelation.
Keywords
blood; cellular biophysics; diseases; haemodynamics; haemorheology; Crit-Line Monitor; GBS; Guillain-Barre Symdrome; Ht values; Tokushima University Hospital; blood pressure; blood purification; hematocrit values; high-dose immunoglobulin therapy; plasma exchange; plasmapheresis; prediction method; red blood cell; Autocorrelation; Biomedical monitoring; Blood pressure; Cells (biology); Hospitals; Medical treatment; Plasma measurements; Prediction methods; Pressure measurement; Purification; Algorithms; Artificial Intelligence; Diagnosis, Computer-Assisted; Guillain-Barre Syndrome; Hematocrit; Humans; Pattern Recognition, Automated; Plasmapheresis; Prognosis; Reproducibility of Results; Sensitivity and Specificity; Therapy, Computer-Assisted;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2007. EMBS 2007. 29th Annual International Conference of the IEEE
Conference_Location
Lyon
ISSN
1557-170X
Print_ISBN
978-1-4244-0787-3
Type
conf
DOI
10.1109/IEMBS.2007.4353000
Filename
4353000