Improved Scalar Transmission Measurements using Vector Source Match Correction
Author
Potter, Chris
Author_Institution
Marconi Instruments, Stevenage, England
Volume
26
fYear
1994
fDate
1-2 Dec. 1994
Firstpage
126
Lastpage
133
Abstract
Scalar transmission measurements are limited mainly by mismatch errors when the insertion loss is small. Unlike vector analyzers, it is not possible to use 12- term error correction to reduce these errors. However, using an inexpensive six - port reflection analyzer and a scalar detector, it is possible to greatly improve measurement performance by correcting for source match error in the transmission measurement. The load match can¿t be errorcorrected, but in practice the source match error dominates since a detector is better matched than the source output or test port of a bridge. This paper shows how a reduced 12- term error model can be used provided the transmission detector is ratioed to the reference detector in the six - port.
Keywords
Bridges; Calibration; Detectors; Error correction; Frequency measurement; Instruments; Loss measurement; Propagation losses; Reflection; Testing;