• DocumentCode
    1842653
  • Title

    Accuracy Improvement of the Automated Multiprobe Transmission Line Reflectometer

  • Author

    L´vov, A.A. ; Semenov, K.V. ; Morzhakov, A.A.

  • Author_Institution
    Saratov State Technical University, Department of Electronic Engineering, 1, Aeroport, Saratov 410019, Russia
  • Volume
    29
  • fYear
    1996
  • fDate
    20-21 June 1996
  • Firstpage
    196
  • Lastpage
    202
  • Abstract
    The method of microwave circuit parameters measurement using the multiprobe transmission line reflectometer (MTLR) is well-known. Measurement process consists in analysis of electromagnetic field distribution inside the line that depends on parameters of an attached load under test (magnitude and phase of reflection coefficient) and absolute value of the MTL incident complex amplitude. Thus, three probes are enough to obtain the system of nodinear equations for unknown parameters. Additional probes are usually used to enlarge the measurement frequency range.
  • Keywords
    Calibration; Cyclic redundancy check; Detectors; Distributed parameter circuits; Electromagnetic measurements; Equations; Microwave measurements; Probes; Transmission line measurements; Transmission lines;