Title :
Continuous fiberoptic intraarterial monitoring of pHa, PaCO2, and PaO2
Author :
Barker, Steven J. ; Tremper, Kevin K. ; Hyatt, John ; Gehrich, John L.
Author_Institution :
California Univ., Irvine Med. Center, Orange, CA, USA
Abstract :
The results of study of a three-component (pH, PaCO2, PaO2) intra-arterial optode in patients undergoing surgical procedures are presented. Twenty-five surgical patients who required radial artery cannulation took part in this study. In each patient, the fiber-optic sensor (0.63 mm in diameter) was inserted into the arterial cannula. Results for the different cannula sizes are analyzed separately. A scatter plot of PoO2 versus PaO2 showing all data from the 18-gauge cannula patients and statistics for the three blood-gas variables and the two cannula sizes are presented and discussed
Keywords :
blood; fibre optics; patient monitoring; surgery; 0.63 mm; 3-component intraarterial optode; PaCO2; PaO2; blood-gas variables; continuous fiberoptic intraarterial monitoring; pHa; radial artery cannulation; scatter plot; surgical patients; DH-HEMTs; Dogs; Monitoring;
Conference_Titel :
Engineering in Medicine and Biology Society, 1989. Images of the Twenty-First Century., Proceedings of the Annual International Conference of the IEEE Engineering in
Conference_Location :
Seattle, WA
DOI :
10.1109/IEMBS.1989.96348