• DocumentCode
    1847780
  • Title

    Processed EMG for closed-loop control of atracurium infusion

  • Author

    Welna, J.O. ; Abenstein, J.P. ; Jones, K.A. ; Lennon, R.L.

  • Author_Institution
    Mayo Clinic, Rochester, MN, USA
  • fYear
    1989
  • fDate
    9-12 Nov 1989
  • Firstpage
    1577
  • Abstract
    The design and application of a closed-loop controller for the infusion of the muscle relaxant atracurium are described. The controller uses a processed electromyography system for the feedback signal and a new control algorithm, for which a flowchart is given. Initial clinical testing in the operating room demonstrated excellent control characteristics. The algorithm settled within 15 min with very little variation in the state of relaxation. Each person had his or her own settling point, but this was always within 7-12% of control, which produced acceptable clinical relaxation
  • Keywords
    biocontrol; bioelectric potentials; closed loop systems; muscle; patient treatment; atracurium infusion; clinical relaxation; clinical testing; closed-loop control; feedback signal; flowchart; muscle relaxant; processed electromyography system; settling point; Anesthesia; Control systems; Drugs; Electromyography; Muscles; Neuromuscular; Patient monitoring; Signal processing; Surgery; System testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1989. Images of the Twenty-First Century., Proceedings of the Annual International Conference of the IEEE Engineering in
  • Conference_Location
    Seattle, WA
  • Type

    conf

  • DOI
    10.1109/IEMBS.1989.96349
  • Filename
    96349