DocumentCode
1849394
Title
Full dual polarization W-band scanning radiometer and stokes parameter measurement
Author
Kim, Sung-Hyun ; Moon, Nam-Won ; Kim, Yong-Hoon
Author_Institution
Sch. of Inf. & Mechatron., Gwangju Inst. of Sci. & Technol., Gwangju, South Korea
fYear
2010
fDate
1-4 March 2010
Firstpage
221
Lastpage
225
Abstract
The W-band fully polarized radiometer with the detail calibration is presented. The calibration standard for the measurement of Stokes parameters was designed and the measured results are presented. For the calibration standard, a wideband analog correlator was designed and described the performance. The correlation coefficient of the U and V Stokes parameter was 0.92 and 0.96, respectively. The detail analysis of the error of the calibration is presented in terms of random and systematic uncertainty.
Keywords
calibration; geophysical equipment; radiometry; Stokes parameters; W-band fully polarized radiometer; calibration standard; correlation coefficient; dual polarization W-band scanning radiometer; wideband analog correlator; Calibration; Measurement uncertainty; Microwave radiometry; Stokes parameters; Temperature measurement; Uncertainty; Stokes parameters; W-band; calibration standard; correlator; polarimetric radiometer;
fLanguage
English
Publisher
ieee
Conference_Titel
Microwave Radiometry and Remote Sensing of the Environment (MicroRad), 2010 11th Specialist Meeting on
Conference_Location
Washington, DC
Print_ISBN
978-1-4244-8120-0
Electronic_ISBN
978-1-4244-8121-7
Type
conf
DOI
10.1109/MICRORAD.2010.5559556
Filename
5559556
Link To Document