• DocumentCode
    1849394
  • Title

    Full dual polarization W-band scanning radiometer and stokes parameter measurement

  • Author

    Kim, Sung-Hyun ; Moon, Nam-Won ; Kim, Yong-Hoon

  • Author_Institution
    Sch. of Inf. & Mechatron., Gwangju Inst. of Sci. & Technol., Gwangju, South Korea
  • fYear
    2010
  • fDate
    1-4 March 2010
  • Firstpage
    221
  • Lastpage
    225
  • Abstract
    The W-band fully polarized radiometer with the detail calibration is presented. The calibration standard for the measurement of Stokes parameters was designed and the measured results are presented. For the calibration standard, a wideband analog correlator was designed and described the performance. The correlation coefficient of the U and V Stokes parameter was 0.92 and 0.96, respectively. The detail analysis of the error of the calibration is presented in terms of random and systematic uncertainty.
  • Keywords
    calibration; geophysical equipment; radiometry; Stokes parameters; W-band fully polarized radiometer; calibration standard; correlation coefficient; dual polarization W-band scanning radiometer; wideband analog correlator; Calibration; Measurement uncertainty; Microwave radiometry; Stokes parameters; Temperature measurement; Uncertainty; Stokes parameters; W-band; calibration standard; correlator; polarimetric radiometer;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Radiometry and Remote Sensing of the Environment (MicroRad), 2010 11th Specialist Meeting on
  • Conference_Location
    Washington, DC
  • Print_ISBN
    978-1-4244-8120-0
  • Electronic_ISBN
    978-1-4244-8121-7
  • Type

    conf

  • DOI
    10.1109/MICRORAD.2010.5559556
  • Filename
    5559556