DocumentCode
1850036
Title
Performance of submillimeter CdZnTe strip detectors
Author
Doty, F.P. ; Butler, J.F. ; Hink, P.L. ; Macri, J.R.
Author_Institution
DIGIRAD, San Diego, CA, USA
Volume
1
fYear
1995
fDate
21-28 Oct 1995
Firstpage
80
Abstract
We report results of a study undertaken to experimentally determine the achievable spatial resolution of monolithic cadmium zinc telluride (CZT) arrays. It was demonstrated that CZT arrays can produce direct electronic X-ray images with fine spatial resolution and high detection efficiency. For low X-ray energies useful for imaging soft tissue such as the breast, 50 micron resolution with virtually 100% detection efficiency were demonstrated, while for higher energy X-rays where fluorescence dominates the signal spreading, 100 micron resolution was demonstrated. Due to electrostatic effects of small anodes, improved charge deficit tailing and pulse rise time results are observed for strips as well as pixels
Keywords
X-ray detection; biomedical imaging; semiconductor counters; 100 micron; 50 micron; CdZnTe; CdZnTe strip detectors; X-ray images; anodes; breast; charge deficit tailing; detection efficiency; low X-ray energies; monolithic; pixels; pulse rise time; soft tissue; spatial resolution; submillimeter; Cadmium compounds; Energy resolution; Image resolution; Signal resolution; Spatial resolution; Strips; X-ray detection; X-ray detectors; X-ray imaging; Zinc compounds;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
Conference_Location
San Francisco, CA
Print_ISBN
0-7803-3180-X
Type
conf
DOI
10.1109/NSSMIC.1995.504181
Filename
504181
Link To Document