• DocumentCode
    1850155
  • Title

    Peak modeling for Ion mobility spectrometry measurements

  • Author

    Kopczynski, Dominik ; Baumbach, Jörg Ingo ; Rahmann, Sven

  • Author_Institution
    Bioinf. for High-Throughput Technol., Tech. Univ. Dortmund, Dortmund, Germany
  • fYear
    2012
  • fDate
    27-31 Aug. 2012
  • Firstpage
    1801
  • Lastpage
    1805
  • Abstract
    Ion mobility spectrometry (IMS), coupled with multicapillary columns (MCCs), is a technology for analyzing the concentration of volatile organic compounds (VOCs) in the air or in exhaled breath. We introduce a new model-based method for peak description and thus for manifold data reduction. Depending on the number of peaks, we reduce the raw data by five orders of magnitude (about 250 000-fold). Each peak is described with seven parameters, which are estimated by an expectation maximization (EM) algorithm that copes well with overlapping peaks. We transform the parameters into robust and interpretable peak descriptors, which are suitable for downstream analysis steps.
  • Keywords
    capillarity; expectation-maximisation algorithm; ion mobility spectrometers; organic compounds; air exhaled breath; downstream analysis; expectation maximization algorithm; inexhaled breath; interpretable peak descriptors; ion mobility spectrometry; manifold data reduction; model-based method; multicapillary columns; overlapping peaks; peak description; volatile organic compounds; Compounds; Computational modeling; Data models; Gaussian distribution; Shape; Spectroscopy; Standards; EM algorithm; Ion mobility spectrometry (IMS); data reduction; mixture model; parameter estimation; three-parameter inverse Gaussian distribution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Signal Processing Conference (EUSIPCO), 2012 Proceedings of the 20th European
  • Conference_Location
    Bucharest
  • ISSN
    2219-5491
  • Print_ISBN
    978-1-4673-1068-0
  • Type

    conf

  • Filename
    6333986