Title :
Performance of a Compton telescope using position-sensitive germanium detectors
Author :
Phlips, B.F. ; Inderhees, S.E. ; Kroeger, R.A. ; Johnson, W.N. ; Kinzer, R.L. ; Kurfess, J.D. ; Graham, B.L. ; Gehrels, N.
Author_Institution :
Univ. Space Res. Assoc., Washington, DC, USA
Abstract :
We describe the use of position sensitive planar germanium detectors in a Compton telescope/camera. This Compton telescope achieves the very good energy resolution (ΔEtotal<4 keV) associated with germanium detectors and good position resolution (2 mm). By combining a 25×25 strip (2 mm pitch) detector with a 5×5 strip detector (9 mm pitch), we created a telescope with 625×25 pixel combinations. Using this detector pair, we have reconstructed positions with an angular resolution of 1 degree FWHM and 15 arcminute centroiding. Point sources are identified with less than 100 full energy events with simple image reconstruction. The angular resolution is currently limited by the uncertainty in the absolute position of the detectors and the size of the second detector pixels. We show the expected angular resolution when the pixel size no longer dominates the angular resolution and discuss proposed applications
Keywords :
astronomical instruments; astronomical telescopes; gamma-ray astronomy; gamma-ray detection; germanium radiation detectors; position sensitive particle detectors; 2 mm; Compton telescope; Ge; angular resolution; energy resolution; pixel size; position resolution; position-sensitive Ge detectors; strip detector; Educational institutions; Energy resolution; Germanium; High-resolution imaging; Image reconstruction; Laboratories; Position sensitive particle detectors; Spatial resolution; Strips; Telescopes;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
Conference_Location :
San Francisco, CA
Print_ISBN :
0-7803-3180-X
DOI :
10.1109/NSSMIC.1995.504202