Title :
Consideration of conditions required for multi-channel simultaneous bioimpedance measurement
Author :
Yamamoto, Y. ; Nakamura, T. ; Kusuhara, T. ; Adli
Author_Institution :
Dept. of Electr. & Electron. Eng., Okayama Univ., Japan
Abstract :
Impedance techniques have been widely applied to the biomedical engineering field. In order to obtain definite results from bioimpedance, multi-channel measurement is effective. A linearity of biological tissue and fundamental technical parameters for the instrument must be confirmed. In this study, the fundamental conditions have been investigated for multi-channel bioimpedance measurement and the condition of measurement for biological tissue has been confirmed. The differences of every measurement frequency should be appropriate to 1 kHz and it has been shown that the linearity of biological tissue is maintained. Based on these results, a two-channel bioimpedance measurement instrument has been constructed
Keywords :
bioelectric phenomena; biomedical measurement; electric impedance measurement; bioimpedance measurement; biological tissue linearity; biomedical engineering; measurement frequency; multi-channel simultaneous measurement; two-channel bioimpedance measurement instrument; Battery charge measurement; Bioimpedance; Biological tissues; Biomedical measurements; Electric variables measurement; Electrical resistance measurement; Electrodes; Frequency measurement; Impedance measurement; Instruments;
Conference_Titel :
Instrumentation and Measurement Technology Conference, 1998. IMTC/98. Conference Proceedings. IEEE
Conference_Location :
St. Paul, MN
Print_ISBN :
0-7803-4797-8
DOI :
10.1109/IMTC.1998.679763