    A 2.4µA continuous-time electrode-skin impedance measurement circuit for motion artifact monitoring in ECG acquisition systems

  • Author

    Kim, Sunyoung ; Yazicioglu, Refet Firat ; Torfs, Tom ; Dilpreet, Buxi ; Julien, Penders ; Van Hoof, Chris

  • Author_Institution
    IMEC, Leuven, Belgium
  • fYear
    2010
  • fDate
    16-18 June 2010
  • Firstpage
    219
  • Lastpage
    220
  • Abstract
    A continuous time electrode skin impedance monitoring system is implemented in parallel with ECG monitoring. To avoid degradation of the ECG signal quality, chopper stabilized AC current sources are adopted. Measured impedance signal is used as a reference signal input for a post processing adaptive filter removing motion artifacts in the ECG signal. The impedance measurement core is implemented in 0.5μm CMOS process and dissipates 2.4μA from a 2V.
  • Keywords
    CMOS integrated circuits; biomedical electrodes; biomedical measurement; electric impedance measurement; electrocardiography; skin; CMOS process; ECG acquisition systems; ECG signal quality; chopper stabilized AC current sources; continuous-time electrode-skin impedance measurement circuit; current 2.4 muA; motion artifact monitoring; reference signal input; size 0.5 mum; voltage 2 V; Choppers; Electrocardiography; Electrodes; Impedance; Impedance measurement; Monitoring; Skin;