Abstract
The ability to monitor and quantify force exertions made by individuals during fMRI scans of the brain would provide researchers and clinicians a standardized, well-controlled behavioral task that could improve the repeatability and accuracy of imaging studies. In this work, we present a MRI compatible wrist module that is capable of measuring isometric forces generated at the hand and joint moments along wrist flexion-extension and wrist ulna-radial deviation axes. Joint moments measured by the system can be visually displayed to the individual and used during target matching tasks. In a small set of pilot tests, it was found that the noise on the force and moment signals were not affected by the magnetic fields nor were the fMRI images influenced by the presence of the device. In future studies, we plan to use the wrist module to investigate cortical reorganization in stroke patients following prolonged neurorehabilitation.
Keywords
biomedical MRI; force sensors; monitoring; patient rehabilitation; real-time systems; MRI compatible force sensing system; fMRI testing; force exertions; neurorehabilitation; real-time monitoring; target matching tasks; wrist moments; Displays; Joints; Magnetic resonance imaging; Monitoring; Optical filters; Optical sensors; Real time systems; System testing; USA Councils; Wrist;
fLanguage
English
Publisher
ieee
Conference_Titel
Rehabilitation Robotics, 2005. ICORR 2005. 9th International Conference on
Print_ISBN
0-7803-9003-2
Type
conf
DOI
10.1109/ICORR.2005.1501087
Filename
1501087
Link To Document