• DocumentCode
    1873030
  • Title

    Flow speed measurement with Doppler effect using ultrasonic receiver for small-sized smart catheter

  • Author

    Matsui, Ryo ; Takei, Yusuke ; Nguyen Minh-Dung ; Takahata, Tomoyuki ; Matsumoto, Kiyoshi ; Shimoyama, Isao

  • Author_Institution
    Univ. of Tokyo, Tokyo, Japan
  • fYear
    2015
  • fDate
    18-22 Jan. 2015
  • Firstpage
    640
  • Lastpage
    642
  • Abstract
    We measured flow speed in a cylindrical pipe with Doppler effect using an ultrasonic receiver. We proposed a wide frequency range receiver to measure both blood flow speed and blood pressure at the same time in a blood vessel. Our sensor has a piezoresistive cantilever which is put on the boundary between air and liquid in a device. According to this structure, proposed receiver can measure acoustic waves from Hz to MHz order frequency, so that we can obtain blood flow speed (MHz order), and blood pressure change (Hz order) with the same device. We measured flow speed ranging from 6 to 25 mm/s, which was equal to blood flow speed of an arteriole, in a cylindrical pipe, of MHz order Doppler effect.
  • Keywords
    Doppler effect; acoustic waves; biomedical ultrasonics; blood flow measurement; blood pressure measurement; blood vessels; catheters; velocity measurement; Doppler effect; acoustic wave; arteriole; blood flow speed measurement; blood pressure measurement; blood vessel; cylindrical pipe; frequency range receiver; piezoresistive cantilever; sensor; small-sized smart catheter; ultrasonic receiver; velocity 25 mm/s; Acoustics; Blood flow; Doppler effect; Fluid flow measurement; Frequency measurement; Receivers; Ultrasonic variables measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Micro Electro Mechanical Systems (MEMS), 2015 28th IEEE International Conference on
  • Conference_Location
    Estoril
  • Type

    conf

  • DOI
    10.1109/MEMSYS.2015.7051037
  • Filename
    7051037