Title :
Elastic registration for stent enhancement in X-ray image sequences
Author :
Bismuth, Vincent ; Vaillant, Régis
Author_Institution :
Gen. Electr. Healthcare, Buc
Abstract :
Curing coronary artery disease typically involves deploying a stent (a fine mesh of wire) supported by a balloon inside the patient artery. This procedure is monitored by X-ray fluoroscopy where stent visibility is challenging. Since 2000, stent enhancement techniques for X-ray fluoroscopy image sequences have become popular. They are based on motion compensated temporal integration of the stent images. The motion of the stent is classically inferred from the motion of the highly contrasted balloon marker balls. We propose a technique that compensates the motion of the whole guide-wire in the neighborhood of the marker balls. We demonstrate that the guide wire-based registration enables better enhanced images in 28% of the cases whereas marker ball based registration is better only in 4% of the cases. We provide examples, qualitative and quantitative analysis on a database of 144 clinical cases.
Keywords :
X-ray imaging; angiocardiography; image enhancement; image registration; image restoration; image sequences; medical image processing; motion compensation; radiography; X-ray fluoroscopy; X-ray image sequences; angiocardiography; coronary artery disease; elastic registration; guide wire-based registration; image restoration; marker ball based registration; motion compensation; stent enhancement; Angioplasty; Arteries; Bismuth; Cardiac disease; Coronary arteriosclerosis; Curing; Image restoration; Image sequences; Medical services; X-ray imaging; Image registration; Image restoration; Image sequence analysis; X-ray angiocardiography;
Conference_Titel :
Image Processing, 2008. ICIP 2008. 15th IEEE International Conference on
Conference_Location :
San Diego, CA
Print_ISBN :
978-1-4244-1765-0
Electronic_ISBN :
1522-4880
DOI :
10.1109/ICIP.2008.4712276