DocumentCode :
18762
Title :
Noninvasive arterial blood pressure waveform monitoring using two- element ultrasound system
Author :
Joohyun Seo ; Pietrangelo, Sabino J. ; Hae-Seung Lee ; Sodini, Charles G.
Author_Institution :
Dept. of Electr. Eng. & Comput. Sci., Massachusetts Inst. of Technol., Cambridge, MA, USA
Volume :
62
Issue :
4
fYear :
2015
fDate :
Apr-15
Firstpage :
776
Lastpage :
784
Abstract :
This work details noninvasive arterial blood pressure (ABP) waveform estimation based on an arterial vessel cross-sectional area measurement combined with an elasticity measurement of the vessel, represented by pulse wave velocity (PWV), using a two-element ultrasound system. The overall ABP waveform estimation is validated in a custom-designed experimental setup mimicking the heart and an arterial vessel segment with two single element transducers, assuming a constant hemodynamic system. The estimation of local PWV using the flow-area method produces unbiased elasticity estimation of the tube in a pressure waveform comparison. The measured PWV using 16 cardiac cycles of data is 8.47 + 0.63 m/s with an associated scaling error of -1.56 + 14.0% in a direct pressure waveform comparison, showing negligible bias error on average. The distension waveform obtained from a complex cross-correlation model estimator (C3M) reliably traces small pressure changes reflected by the diameter change. The excellent agreement of an estimated pressure waveform to the reference pressure waveform suggests the promising potential of a readily available, inexpensive, and portable ABP waveform monitoring device.
Keywords :
biomedical transducers; blood pressure measurement; blood vessels; elasticity; patient monitoring; ultrasonic transducers; ABP waveform estimation; PWV; arterial vessel cross-sectional area measurement; cross-correlation model estimator; distension waveform; elasticity measurement; estimated monitoring device; flow-area method; heart; hemodynamic system; noninvasive arterial blood pressure waveform estimation; noninvasive arterial blood pressure waveform monitoring; pulse wave velocity; single element transducers; two-element ultrasound system; waveform pressure cardiac cycles; Area measurement; Electron tubes; Estimation; Pressure measurement; Transducers; Ultrasonic imaging; Ultrasonic variables measurement;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2014.006904
Filename :
7081473
Link To Document :