Title :
High Accuracy Measurement System of Absorbed Dose using Combination of Radiochromic Film and Micro-Densitometer for Radiation Therapy
Author :
Aoyama, Hideki ; Azuma, Yoshiharu ; Goto, Sachiko ; Hato, Hirofumi ; Sumimoto, Tetsuhiro
Author_Institution :
Central Div. of Radiol., Okayama Univ. Hosp.
Abstract :
Radiochromic film (RCF) has been applied to mapping of dose distributions in radiation therapy. However, this particular RCF product suffers from intrinsic nonuniform response. In this study, we have carefully measured the nonuniformity of unirradiated RCF, and the response of irradiated RCF in pixel-to-pixel as function of various absorbed doses, in order to ascertains for the method of predicting the error of irradiated, unirradiated RCF
Keywords :
densitometry; dosimetry; radiation therapy; dose distributions mapping; dose measurement system; microdensitometer; nonuniform response; radiation dosimetry; radiation therapy; radiochromic film; Biomedical applications of radiation; Dosimetry; Energy measurement; Instrumentation and measurement; Optical sensors; Radiation detectors; Radiology; Semiconductor films; Semiconductor radiation detectors; Spatial resolution; Nonuniform response; Radiation dosimetry; Radiochromic film;
Conference_Titel :
Instrumentation and Measurement Technology Conference, 2006. IMTC 2006. Proceedings of the IEEE
Conference_Location :
Sorrento
Print_ISBN :
0-7803-9359-7
Electronic_ISBN :
1091-5281
DOI :
10.1109/IMTC.2006.328504