• DocumentCode
    1879085
  • Title

    High Accuracy Measurement System of Absorbed Dose using Combination of Radiochromic Film and Micro-Densitometer for Radiation Therapy

  • Author

    Aoyama, Hideki ; Azuma, Yoshiharu ; Goto, Sachiko ; Hato, Hirofumi ; Sumimoto, Tetsuhiro

  • Author_Institution
    Central Div. of Radiol., Okayama Univ. Hosp.
  • fYear
    2006
  • fDate
    24-27 April 2006
  • Firstpage
    403
  • Lastpage
    408
  • Abstract
    Radiochromic film (RCF) has been applied to mapping of dose distributions in radiation therapy. However, this particular RCF product suffers from intrinsic nonuniform response. In this study, we have carefully measured the nonuniformity of unirradiated RCF, and the response of irradiated RCF in pixel-to-pixel as function of various absorbed doses, in order to ascertains for the method of predicting the error of irradiated, unirradiated RCF
  • Keywords
    densitometry; dosimetry; radiation therapy; dose distributions mapping; dose measurement system; microdensitometer; nonuniform response; radiation dosimetry; radiation therapy; radiochromic film; Biomedical applications of radiation; Dosimetry; Energy measurement; Instrumentation and measurement; Optical sensors; Radiation detectors; Radiology; Semiconductor films; Semiconductor radiation detectors; Spatial resolution; Nonuniform response; Radiation dosimetry; Radiochromic film;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2006. IMTC 2006. Proceedings of the IEEE
  • Conference_Location
    Sorrento
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-9359-7
  • Electronic_ISBN
    1091-5281
  • Type

    conf

  • DOI
    10.1109/IMTC.2006.328504
  • Filename