• DocumentCode
    1880533
  • Title

    Dynamic EIT image reconstruction using regularization based on conjugate gradients

  • Author

    Yan, Peimin ; Yuan, Zhi

  • Author_Institution
    Sch. of Commun. & Inf. Eng., Shanghai Univ., Shanghai
  • fYear
    2008
  • fDate
    25-28 Oct. 2008
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    In order to solving the illness problem and improving the quality of the reconstructed image in dynamic Electrical Impedance Tomography (EIT), this paper proposes a method of image reconstruction using the conjugate gradients (CG) algorithm with regularized pre-processing factors. As the number of condition in coefficient matrix is decreased, the algorithm will have more rapid convergence. The imaging result of computer simulation and salt water tank experiment has shown that when the impedance change is small in dynamic EIT imaging, certain accuracy can be guaranteed in EIT result approximation using this method.
  • Keywords
    conjugate gradient methods; electric impedance imaging; medical image processing; conjugate gradients algorithm; dynamic EIT image reconstruction; electrical impedance tomography; illness problem; salt water tank experiment; Approximation algorithms; Character generation; Computer simulation; Convergence; Image reconstruction; Impedance; Large-scale systems; Linear approximation; Linear systems; Tomography; conjugate gradients; image reconstruction; regularization;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Applied Sciences on Biomedical and Communication Technologies, 2008. ISABEL '08. First International Symposium on
  • Conference_Location
    Aalborg
  • Print_ISBN
    978-1-4244-2647-8
  • Electronic_ISBN
    978-1-4244-2648-5
  • Type

    conf

  • DOI
    10.1109/ISABEL.2008.4712613
  • Filename