Abstract :
The uncertainty of transit time ultrasonic clamp-on flowmeters is relatively high (currently approximately plusmn5% (Sanderson et al., 2002)). We have studied several factors contributing to this uncertainty, focusing mainly on the separation distance between the transducers. This paper will discuss experimental work and the use of ray tracing and finite element models to examine this relationship.
Keywords :
finite element analysis; flowmeters; measurement uncertainty; ray tracing; ultrasonic transducers; clamp-on transit time ultrasonic flowmeter uncertainty; finite element model; ray tracing; transducers; Equations; Finite element methods; Fluid flow measurement; Instrumentation and measurement; Liquids; Ray tracing; Transmitters; Ultrasonic transducers; Ultrasonic variables measurement; Uncertainty; clamp-on; flow measurement; flowmeter; ray model; ultrasonic;