• DocumentCode
    191186
  • Title

    Broadband dielectric spectroscopy calibration for microliter samples of biogenic liquid

  • Author

    Liu, S. ; Ocket, I. ; Barmuta, P. ; Lewandowski, A. ; Schreurs, D. ; Nauwelaers, B.

  • Author_Institution
    Div. ESAT-TELEMIC, Univ. of Leuven, Leuven, Belgium
  • fYear
    2014
  • fDate
    6-9 Oct. 2014
  • Firstpage
    279
  • Lastpage
    282
  • Abstract
    This paper reports on dielectric spectroscopy for microliter liquid samples using reference materials. In order to fully determine the calibration coefficients, air and water are used as known reference materials, while methanol is used as an unknown material, which is also used to validate the approach up to 5 GHz for which reference data is available. The method is then used up to 110 GHz for yeast cell suspensions and beef protein solutions.
  • Keywords
    bioelectric phenomena; cellular biophysics; molecular biophysics; proteins; suspensions; beef protein solutions; biogenic liquid; broadband dielectric spectroscopy; calibration coefficients; methanol; microliter liquid samples; yeast cell suspensions; Calibration; Coplanar waveguides; Liquids; Permittivity measurement; Proteins; Sensors; Transmission line measurements; biology; calibration; dielectrics; millimeter wave; on-wafer; spectroscopy;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference (EuMC), 2014 44th European
  • Conference_Location
    Rome
  • Type

    conf

  • DOI
    10.1109/EuMC.2014.6986424
  • Filename