• DocumentCode
    1919374
  • Title

    Non-invasive measurement of blood sugar level by millimeter waves

  • Author

    Nikawa, Y. ; Someya, D.

  • Author_Institution
    Dept. of Electr. Eng., Kokushikan Univ., Tokyo, Japan
  • Volume
    1
  • fYear
    2001
  • fDate
    20-24 May 2001
  • Firstpage
    171
  • Abstract
    Measurement of blood sugar level by noninvasive method is highly desirable. Nevertheless, a practical non-invasive technique has not been reported. In this paper, to measure blood sugar level using millimeter wave noninvasively, complex permittivity of glucose and sodium chloride solution is measured and the change of transmission coefficient in millimeter wave region is obtained. An applicator system to measure blood sugar level in vivo is designed and optimized by simulation using transmission line modeling (TLM) method.
  • Keywords
    biomedical measurement; blood; millimetre wave measurement; permittivity; applicator system; blood sugar level; complex permittivity; millimeter wave applications; noninvasive measurement; transmission coefficient; transmission line modeling; Blood; Diabetes; Frequency; Humans; In vivo; Millimeter wave measurements; Millimeter wave technology; Permittivity measurement; Sugar; Transmission line measurements;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Symposium Digest, 2001 IEEE MTT-S International
  • Conference_Location
    Phoenix, AZ, USA
  • ISSN
    0149-645X
  • Print_ISBN
    0-7803-6538-0
  • Type

    conf

  • DOI
    10.1109/MWSYM.2001.966865
  • Filename
    966865