Evaluation of cerium doped lutetium oxyorthosilicate (LSO) scintillation crystals for PET
Author :
Daghighiam, F. ; Shenderov, P. ; Pentlow, K.S. ; Graham, M.C. ; Melcher, C.L. ; Schweitzer, J.S.
Author_Institution :
Memorial Sloan-Kettering Cancer Center, New York, NY, USA
fYear :
1992
fDate :
25-31 Oct 1992
Firstpage :
876
Abstract :
A new scintillation crystal, cerium-doped lutetium oxyorthosilicate (LSO), was recently discovered, with light intensity 75% of NaI(Tl), scintillation decay time of 12 ns (30%) and 42 ns (70%), effective Z of 66, and density of 7.4 g/cc. The fast decay time and scintillation light output of LSO are superior to those of bismuth germanate (BGO) for positron emission tomography (PET) and the stopping power of LSO for 511 keV photons is only slightly lower than that of BGO. The authors directly compared the detection characteristics relevant to PET applications of small crystals of LSO (2×2×10 mm) with those of BGO. The energy resolution at 511 keV was 12% FWHM for LSO and two to three times wider for BGO. The coincidence timing of two opposing crystals, using a position sensitive photomultiplier tube (PMT) (Hamamatsu R2486), were 1.4 ns FWHM for LSO and more than three times higher for BGO. Using a fast PMT (Hamamatsu R3177), coincidence timing for LSO yielded 0.46 ns FWHM. These crystals are being used to simulate a small-scale PET scanner and to investigate its imaging performance
Keywords :
biomedical equipment; computerised tomography; radioisotope scanning and imaging; 0.46 ns; 1.4 ns; 12 ns; 42 ns; 511 keV; Ce-doped Lu oxyorthosilicate; NaI:Tl; coincidence timing; decay time; detection characteristics; medical diagnostic imaging; medical instrumentation; nuclear medicine; position sensitive photomultiplier tube; positron emission tomography; scintillation crystals; scintillation light output; Cerium; Energy resolution; Optical attenuators; Optical imaging; Optical scattering; Optical sensors; Photonic crystals; Positron emission tomography; Solid scintillation detectors; Timing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
0-7803-0884-0
Type :
conf
DOI :
10.1109/NSSMIC.1992.301022
Filename :
301022
Link To Document :
بازگشت