DocumentCode :
1924209
Title :
Study of reprojection methods in terms of their resolution loss and sampling errors
Author :
Yu, Dan-Chu ; Huang, Sung-Cheng
Author_Institution :
California Univ., Sch. of Med., Los Angeles, CA, USA
fYear :
1992
fDate :
25-31 Oct 1992
Firstpage :
1160
Abstract :
Two reprojection methods were studied with respect to their resolution loss and sampling errors. The modulation transfer function was used to evaluate the resolution loss, and the residual sum of squares (RSS) of the difference between the reprojected sinogram and the reference sinogram was used to assess the sampling errors. The two methods studied are: area weighted convolution, (AWC) and Gaussian pixel convolution (GPC). For both methods, resolution loss was determined by the reconstruction filter, the linear interpolation in the backprojection, and the convolution kernel. Sampling errors were found to be angle-dependent, and the angular dependence was more pronounced for the GPC. To avoid significant sampling error, the width of the convolution kernel needs to be two times larger than the pixel distance. The large subbinning size of the projection array lead to interpolation error, which is more pronounced for AWC than for GPC. With pixel distance equal to half the sampling distance, the sampling error can be greatly reduced without additional resolution loss
Keywords :
computerised tomography; image reconstruction; medical image processing; optical transfer function; Gaussian pixel convolution; angular dependence; area weighted convolution; backprojection; convolution kernel; filtered backprojection image reconstruction; interpolation error; modulation transfer function; pixel distance; reprojected sinogram; reprojection methods; residual sum of squares; resolution loss; sampling errors; subbinning size; Convolution; Fourier transforms; Identity-based encryption; Image reconstruction; Image sampling; Interpolation; Kernel; Nonlinear filters; Nuclear medicine; Sampling methods;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
0-7803-0884-0
Type :
conf
DOI :
10.1109/NSSMIC.1992.301048
Filename :
301048
Link To Document :
بازگشت