DocumentCode :
1925338
Title :
Effect of lead filters on the performance of a neuro-PET tomograph operated without septa
Author :
Spinks, T.J. ; Shah, S.I.
Author_Institution :
Hammersmith Hospital, London,UK
fYear :
1992
fDate :
25-31 Oct 1992
Firstpage :
939
Abstract :
The value of using lead filters to improve the quality of PET (positron emission tomography) data has been investigated for a CTI/Siemens 953B brain scanner operated without septa. The filters consisted of sheets of thickness 0.5 mm and 1.0 mm and axial width 18 cm, which were placed around the outer surface of the patient port. The port diameter is 35 cm and the axial FO (field of view) is 10.65 cm. The ratios of dead times (with/without filter) with 1 mCi (20 cm uniform cylinder) in the FOV are 0.92 (0.5 mm) and 0.82 (1 mm) at a lower threshold (LLD) of 380 keV; the values at 150 keV are 0.87 and 0.76. Random rates are reduced by 12% (0.5 mm) and 27% (1 mm) at a LLD of 380 keV and by 20% and 40% at 150 keV. These values reflect the significant reduction of lower energy events observed in the measured energy spectrum. However, these reductions are accompanied by a drop in the true coincidence rate such that noise-equivalent counts are reduced for count rates up to the maximum measured trues rate at lower thresholds of 380 keV and 150 keV. It is concluded that the filters used in this study provide no advantage in the performance of the tomograph
Keywords :
biomedical equipment; brain; computerised tomography; radioisotope scanning and imaging; 10-3 Ci; 150 keV; 20 cm; 35 cm; 380 keV; CTI/Siemens 953B brain scanner; Pb filters effects; axial width; dead time; energy spectrum; medical diagnostic imaging; neuro-PET tomograph; noise-equivalent counts; nuclear medicine; patient port; positron emission tomography; septa; true coincidence rate; uniform cylinder; Brain; Cameras; Cyclotrons; Detectors; Energy measurement; Face detection; Filters; Hospitals; Noise reduction; Positron emission tomography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
0-7803-0884-0
Type :
conf
DOI :
10.1109/NSSMIC.1992.301100
Filename :
301100
Link To Document :
بازگشت