Title :
A ring Compton scatter camera for imaging medium energy gamma rays
Author :
Martin, J.B. ; Knoll, G.F. ; Wehe, D.K. ; Dogan, N. ; Jordanov, V. ; Petrick, N. ; Singh, M.
Author_Institution :
Dept. of Nucl. Eng., Michigan Univ., Ann Arbor, MI, USA
Abstract :
Characteristics of a novel ring Compton scatter camera are described. This camera is designed to image medium-energy (0.5 to 3 MeV) gamma-ray fields. It consists of two position-sensitive detector arrays: a 4×4 planar array of high purity germanium (HPGe) crystals and a ring array of up to 64 NaI(Tl) crystals. Past evaluations of Compton cameras have used a planar second detector, which is subjected to a large flux of gamma rays that either pass directly through or undergo small angle scatter in the first detector. A ring array significantly reduces direct and small angle scattered events in the second detector. An analytical model for ring camera systems has been developed to predict angular resolution and efficiency, and has been benchmarked against measurements made with a prototype system consisting of the 4×4 HPGe array and an eight-element ring. Predictions are made for a system with 64 crystals in the second detector ring
Keywords :
gamma-ray detection and measurement; position sensitive particle detectors; scintillation counters; semiconductor counters; 0.5 to 3 MeV; Ge detector; HPGe detector; NaI(Tl) crystals; NaI:Tl; medium energy gamma rays; position-sensitive detector; ring Compton scatter camera; Cameras; Crystals; Gamma ray detection; Gamma ray detectors; Gamma rays; Optical imaging; Planar arrays; Position sensitive particle detectors; Scattering; Sensor arrays;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
0-7803-0884-0
DOI :
10.1109/NSSMIC.1992.301187