DocumentCode
1931344
Title
Development of new cancer treatment using approved MRI contrast agent and induction heating device
Author
Nagae, Hideo ; Nagano, Isamu ; Yagitani, Satoshi ; Yamada, Sotoshi ; Katayama, Kanji ; Tazawa, Kenji ; Ikehata, Yoshio
Author_Institution
Kanazawa Univ., Kanazawa, Japan
fYear
2011
fDate
13-20 Aug. 2011
Firstpage
1
Lastpage
4
Abstract
MRI contrast agents, Feridex and Resovist have superparamagnetic substance as an active ingredients. Feridex does not show temperature rise under AC magnetic field. In contrast Resovist shows temperature rise under the same condition. We simulated the optimum frequency on Resovist using phantom and found it is 140 kHz. However treatable cancer volume was limited to 9cm3. We tried a fractionation of Resovist. One of the fractions was found to be effective by more than 6 times compared to the drug by itself. This result suggests that the hyperthermia in combination with the fraction and induction heating device is a promising treatment for cancer.
Keywords
biomedical MRI; cancer; hyperthermia; phantoms; radiation therapy; Feridex; MRI contrast agent; Resovist; cancer treatment; frequency 140 kHz; hyperthermia; induction heating device; phantom; Cancer; Drugs; Electromagnetic heating; Hyperthermia; Magnetic resonance imaging; Rabbits; Tumors;
fLanguage
English
Publisher
ieee
Conference_Titel
General Assembly and Scientific Symposium, 2011 XXXth URSI
Conference_Location
Istanbul
Print_ISBN
978-1-4244-5117-3
Type
conf
DOI
10.1109/URSIGASS.2011.6051352
Filename
6051352
Link To Document