• DocumentCode
    1931905
  • Title

    Calibration with single measurement in microwave imaging system for breast cancer detection

  • Author

    Sollip Kwon ; Yu Heesun ; Seungjun Lee

  • Author_Institution
    Dept. of Electron. Eng., Ewha Womans Univ., Seoul, South Korea
  • fYear
    2015
  • fDate
    13-17 April 2015
  • Firstpage
    1
  • Lastpage
    2
  • Abstract
    Microwave imaging systems for breast cancer detection go through calibration process to extract tumor response from measured data. All the calibration methods that have been proposed so far require multiple measurements. In this paper, we propose a novel calibration method that requires only single measurement. Simulation results with 3D hemispherical breast model show that the restored image is better than that obtained with conventional array rotate calibration in terms of signal to mean ratio, signal to clutter ratio, and localization error.
  • Keywords
    calibration; cancer; image restoration; medical image processing; microwave imaging; microwave measurement; tumours; 3D hemispherical breast model; breast cancer detection; calibration process; conventional array rotate calibration; localization error; restored image; signal-to-clutter ratio; signal-to-mean ratio; single microwave imaging system measurement; tumor response; Arrays; Breast; Calibration; Microwave imaging; Microwave measurement; Microwave theory and techniques; Tumors; breast cancer; calibration; microwave imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Antennas and Propagation (EuCAP), 2015 9th European Conference on
  • Conference_Location
    Lisbon
  • Type

    conf

  • Filename
    7228690