Title :
Development of Precise-temperature-controlled Cooling Apparatus for Medical Application by Using Peltier Effect
Author :
Maruyama, Shigenao ; Komiya, Atsuki ; Takeda, Hiroki ; Aiba, Setsuya
Author_Institution :
Inst. of Fluid Sci., Tohoku Univ., Tokyo
Abstract :
This paper describes precise temperature controlled cooling apparatuses for medical application. A Peltier cryoprobe for dermis treatment and flexible cryoprobe for cryosurgery for deep region of human body were developed by utilizing the Peltier effect. The evaluation experiment of the Peltier cryoprobe was conducted. The proposed apparatus has enough performance to necrotize target diseased cells precisely. The proposed flexible cryoprobe is composed of a flexible resin tube enough to treat the deeper parts of the human body with a catheter. A Peltier element is attached at the top of the probe for active heat-transfer control of the cooling tip. The cooling experiments are carried out and the performance of the cryoprobe is evaluated. The achieved minimum temperature at the tip is -43degC.
Keywords :
Peltier effect; biomedical engineering; biothermics; cooling; patient treatment; probes; Peltier cryoprobe; Peltier effect; cryosurgery; dermis treatment; flexible resin tube; precise temperature controlled cooling apparatus; temperature 255 K; Biomedical equipment; Catheters; Cooling; Dermis; Humans; Medical services; Probes; Resins; Temperature control; Thermoelectricity;
Conference_Titel :
BioMedical Engineering and Informatics, 2008. BMEI 2008. International Conference on
Conference_Location :
Sanya
Print_ISBN :
978-0-7695-3118-2
DOI :
10.1109/BMEI.2008.239