DocumentCode
1939369
Title
Evaluation of Hematocrit Measurement Using Spectral Domain Optical Coherence Tomography
Author
Xu, Xiangqun ; Chen, Zhongping
Author_Institution
Sch. of Sci., Zhejiang Sci-Tech Univ., Hangzhou
Volume
2
fYear
2008
fDate
27-30 May 2008
Firstpage
615
Lastpage
618
Abstract
Optical coherence tomography (OCT) has the potential to be a noninvasive method for hematocrit (HCT) measurement. This paper studied the effect of erythrocyte aggregation on hematocrit measurement based on monitoring the slope changes in OCT depth reflectivity profile using spectral domain optical coherence tomography (SDOCT). Our measurement indicates that the HCT estimated by SDOCT depends on erythrocyte aggregation state. Measured hematocrit in the blood samples with 2%Dextran500 is underestimated by 17.1-19.5% for HCT from 35- 55% at a flow rate of 4.7 mm/s. The results indicate erythrocyte aggregation is an important factor that will affect accurate hematocrit measurement using SDOCT. Developing methods of simultaneous measurements of BBC aggregation and hematocrit is essential for accurate determination of HCT.
Keywords
biomedical measurement; biomedical optical imaging; blood; cellular biophysics; optical tomography; BBC; Dextran 500; OCT depth reflectivity profile; blood; erythrocyte aggregation; hematocrit measurement; slope changes; spectral domain optical coherence tomography; Biomedical measurements; Biomedical optical imaging; Humans; Optical attenuators; Optical refraction; Optical scattering; Optical variables control; Red blood cells; Reflectivity; Tomography; erythrocyte aggregation; hematocrit; optical coherence tomography;
fLanguage
English
Publisher
ieee
Conference_Titel
BioMedical Engineering and Informatics, 2008. BMEI 2008. International Conference on
Conference_Location
Sanya
Print_ISBN
978-0-7695-3118-2
Type
conf
DOI
10.1109/BMEI.2008.280
Filename
4549248
Link To Document