• DocumentCode
    1954361
  • Title

    Simulating capacitive micromachined ultrasonic transducers (CMUTs) using field II

  • Author

    Bæk, David ; Oralkan, Ömer ; Kupnik, Mario ; Willatzen, Morten ; Khuri-Yakub, Butrus T. ; Jensen, Jørgen Arendt

  • Author_Institution
    Dept. of Electr. Eng., Tech. Univ. of Denmark, Lyngby, Denmark
  • fYear
    2010
  • fDate
    11-14 Oct. 2010
  • Firstpage
    439
  • Lastpage
    442
  • Abstract
    Field II has been a recognized simulation tool for piezoceramic medical transducer arrays for more than a decade. The program has its strength in doing fast computations of the spatial impulse response (SIR) from array elements by dividing the elements into smaller mathematical elements (ME)s from which it calculates the SIR responses. The program features predefined models for classical transducer geometries, but currently none for the fast advancing CMUTs. This work addresses the assumptions required for modeling CMUTs with Field II. It is shown that rectangular array elements, populated with cells, can be well approximated by neglecting the cells. Further, it is demonstrated that scaling of the SIR translates into better computational efficiency.
  • Keywords
    acoustic field; biomedical transducers; biomedical ultrasonics; capacitive sensors; micromachining; piezoceramics; piezoelectric transducers; ultrasonic transducer arrays; capacitive micromachined ultrasonic transducers; piezoceramic medical transducer arrays; rectangular array elements; spatial impulse response; ultrasonic field II; Acoustics; Biomedical imaging; Geometry; Pistons; Transducers; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2010 IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4577-0382-9
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2010.5935580
  • Filename
    5935580