Abstract :
This paper describes the design, fabrication, and experimental results of a novel transducer for measuring intravascular blood flow. The transducer consists of a diffraction-grating transducer, to provide vector Doppler capability, wrapped around a cylinder for mounting on a catheter. The new device appears to be useful for quantitative flow measurements to guide and assess intravascular procedures.
Keywords :
biomedical ultrasonics; blood flow measurement; catheters; ultrasonic transducers; catheter; helical diffraction grating transducer; intravascular blood flow measurement; vector Doppler capability; Catheters; Doppler effect; Electrodes; Fluid flow measurement; Transducers; Ultrasonic imaging; Velocity measurement; Doppler; Vector Doppler; blood flow; instrumentation;