• DocumentCode
    1963737
  • Title

    Development of an accurate ultrasonic measurement system using an optical interferometer

  • Author

    Matsuda, Youichi ; Yoshioka, Masahiro ; Uchida, Takeyoshi ; Kikuchi, Tsuneo

  • Author_Institution
    AIST, Nat. Metrol. Inst. of Japan, Tsukuba, Japan
  • fYear
    2010
  • fDate
    11-14 Oct. 2010
  • Firstpage
    273
  • Lastpage
    276
  • Abstract
    An ultrasonic measurement system based on the reflection method using a combination of an optical interferometer and an air-backed pellicle was developed and validated. The measured effective refractive index of water for far-field ultrasounds was roughly equivalent to the theoretical value. The acousto-optic effect in focused ultrasound was also investigated using the system.
  • Keywords
    acousto-optical effects; light interferometry; refractive index measurement; ultrasonic focusing; ultrasonic measurement; ultrasonic reflection; ultrasonic refraction; ultrasonic transducers; ultrasonic variables measurement; acoustooptic effect; air-backed pellicle; effective refractive index; far-field ultrasound; focused ultrasound; optical interferometer; reflection method; ultrasonic measurement system; water; Acoustics; Frequency measurement; Optical interferometry; Refractive index; Temperature measurement; Ultrasonic imaging; Ultrasonic variables measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2010 IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4577-0382-9
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2010.5935991
  • Filename
    5935991