    In vivo photoacoustic and pulse echo imaging of a pancreatic tumor using a hand held device

  • Author

    Montilla, Leonardo G. ; Olafsson, Ragnar ; Witte, Russell S.

  • Author_Institution
    Coll. of Opt. Sci., Univ. of Arizona, Tucson, AZ, USA
  • fYear
    2010
  • fDate
    11-14 Oct. 2010
  • Firstpage
    2147
  • Lastpage
    2150
  • Abstract
    Ultrasonography and computed tomography are often used to diagnose pancreatic cancer. Using similar equipment as ultrasonography, photoacoustic (PA) imaging can provide vascular information over the same region of interest. Information about the vascularity in and around a lesion can be used to aid in the characterization and diagnosis of various cancers. Current PA imaging setups are restricted to bulky bench-top setups, limiting its practicality for clinical research. An ideal imaging platform would be non-invasive, real-time, portable, and inexpensive. We designed and fabricated an attachment to a clinical ultrasound probe which houses an optically transparent acoustic reflector in water. The design enables laser illumination in-line with the acoustic propagation path. The detector array simultaneously captures unbeamformed data on 64 elements at frequencies up to 10 MHz. We used the device to image in vivo a subcutaneous tumor in a SCID mouse implanted with Capan-2 pancreatic cancer cells and also a mouse pancreas embedded in a gel. Laser pulses (5 ns, 13 mJ/cm2) were transmitted through an optical window in the device, providing line illumination below the skin surface. Real time 2D PA images between 700 nm and 960 nm were captured along with conventional pulse echo (PE) ultrasound to examine different sources of contrast in the tumor. The PE image identifies the acoustic window, skin surface, and variation in acoustic impedance within the tumor. PA images visualized areas of near infrared light absorption 4 mm deep within the tumor. Co-registered PE images provided an anatomical reference. This in vivo study demonstrates how a simple adapter to a clinical ultrasound array can be used for real-time simultaneous PE and PA imaging of cancer. With this efficient and practical design, cancer research can capitalize on noninvasive optical contrast below the tissue surface. With an optimized design, clinicians can incorporate PA imaging with routine ultras- - ound exams.
  • Keywords
    biomedical optical imaging; biomedical ultrasonics; cancer; data analysis; image registration; laser applications in medicine; medical image processing; photoacoustic effect; skin; tumours; Capan-2 pancreatic cancer cells; PA imaging; SCID mouse; acoustic impedance; acoustic propagation; clinical ultrasound array; clinical ultrasound probe; coregistered PE images; detector array; hand held device; in-vivo photoacoustic imaging; in-vivo pulse echo imaging; laser illumination; lesion; mouse pancreas; near infrared light absorption; noninvasive optical contrast; optical window; optically transparent acoustic reflector; pancreatic tumor; real-time 2D PA images; skin surface; subcutaneous tumor; tissue surface; unbeamformed data; vascular information; Cancer; Mice; Optical imaging; Probes; Tumors; Ultrasonic imaging; clinical ultrasound; in-line illumination; multi-modality imaging; pancreatic cancer imaging; photoacoustic spectroscopy; ultrasonography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2010 IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4577-0382-9
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2010.5936016
  • Filename
    5936016