Title :
Design and implementation of a system for measuring pupil reactivity using computer vision
Author :
Pérez, Jaime G Barrero ; Gómez, Arturo Plata ; Díaz, Juan Camilo Avendaño ; Larrotta, Fabián Núñez
Author_Institution :
Grupo de Investig. en Control, Electron., Modelado y Simulacion, Univ. Ind. de Santander, Bucaramanga, Colombia
Abstract :
A pupillometric system was developed to be used at the Neuroscience and Behavior Research Group from Universidad Industrial de Santander. First of all, an acquisition system (camera) was selected and implemented, taking into account that pupil diameter varies from 1.5 mm up to 8 or 9 mm, reacting to stimuli in 0.2 seconds. Then, the selected camera was characterized, a measurement and calibration protocol was implemented, and an algorithm for measuring pupil diameter was established. The system gives pupil diameter as a function of time, with an accuracy of ± 0.2 mm.
Keywords :
cameras; computer vision; medical image processing; virtual instrumentation; LabVIEW; calibration protocol; camera acquisition system; computer vision; measuring pupil reactivity design; pupillometric system; time 0.2 s; Biomedical imaging; Electronic mail; Image segmentation; Light emitting diodes; Signal processing; Software; Videos; Computer Vision; LabVIEW; Near Infrared; Pupillometry;
Conference_Titel :
Image, Signal Processing, and Artificial Vision (STSIVA), 2012 XVII Symposium of
Conference_Location :
Antioquia
Print_ISBN :
978-1-4673-2759-6
DOI :
10.1109/STSIVA.2012.6340565