DocumentCode
1972124
Title
Design and implementation of a system for measuring pupil reactivity using computer vision
Author
Pérez, Jaime G Barrero ; Gómez, Arturo Plata ; Díaz, Juan Camilo Avendaño ; Larrotta, Fabián Núñez
Author_Institution
Grupo de Investig. en Control, Electron., Modelado y Simulacion, Univ. Ind. de Santander, Bucaramanga, Colombia
fYear
2012
fDate
12-14 Sept. 2012
Firstpage
104
Lastpage
109
Abstract
A pupillometric system was developed to be used at the Neuroscience and Behavior Research Group from Universidad Industrial de Santander. First of all, an acquisition system (camera) was selected and implemented, taking into account that pupil diameter varies from 1.5 mm up to 8 or 9 mm, reacting to stimuli in 0.2 seconds. Then, the selected camera was characterized, a measurement and calibration protocol was implemented, and an algorithm for measuring pupil diameter was established. The system gives pupil diameter as a function of time, with an accuracy of ± 0.2 mm.
Keywords
cameras; computer vision; medical image processing; virtual instrumentation; LabVIEW; calibration protocol; camera acquisition system; computer vision; measuring pupil reactivity design; pupillometric system; time 0.2 s; Biomedical imaging; Electronic mail; Image segmentation; Light emitting diodes; Signal processing; Software; Videos; Computer Vision; LabVIEW; Near Infrared; Pupillometry;
fLanguage
English
Publisher
ieee
Conference_Titel
Image, Signal Processing, and Artificial Vision (STSIVA), 2012 XVII Symposium of
Conference_Location
Antioquia
Print_ISBN
978-1-4673-2759-6
Type
conf
DOI
10.1109/STSIVA.2012.6340565
Filename
6340565
Link To Document