  • Abstract
    A pupillometric system was developed to be used at the Neuroscience and Behavior Research Group from Universidad Industrial de Santander. First of all, an acquisition system (camera) was selected and implemented, taking into account that pupil diameter varies from 1.5 mm up to 8 or 9 mm, reacting to stimuli in 0.2 seconds. Then, the selected camera was characterized, a measurement and calibration protocol was implemented, and an algorithm for measuring pupil diameter was established. The system gives pupil diameter as a function of time, with an accuracy of ± 0.2 mm.
  • Keywords
    cameras; computer vision; medical image processing; virtual instrumentation; LabVIEW; calibration protocol; camera acquisition system; computer vision; measuring pupil reactivity design; pupillometric system; time 0.2 s; Biomedical imaging; Electronic mail; Image segmentation; Light emitting diodes; Signal processing; Software; Videos; Computer Vision; LabVIEW; Near Infrared; Pupillometry;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Image, Signal Processing, and Artificial Vision (STSIVA), 2012 XVII Symposium of
  • Conference_Location
    Antioquia
  • Print_ISBN
    978-1-4673-2759-6
  • Type

    conf

  • DOI
    10.1109/STSIVA.2012.6340565
  • Filename
    6340565