DocumentCode :
19817
Title :
Edge localisation in MRI for images with signal-dependent noise
Author :
Aarya, I. ; Jiang, D. ; Gale, T.
Author_Institution :
Sch. of Eng. & ICT, Univ. of Tasmania, Hobart, TAS, Australia
Volume :
51
Issue :
15
fYear :
2015
fDate :
7 23 2015
Firstpage :
1151
Lastpage :
1153
Abstract :
A new edge localisation technique for step edges in images with signal-dependent Rician noise in MRI is proposed. Dependent noise can not only affect the detection of true edges in an image but also their position. Inaccurate localisation of an edge can lead to insufficient segmentation and reduce the overall accuracy of any edge detection method. The proposed technique uses higher-order moments of the noise function to determine the correction factor for localisation and thus reduce the overall mean square error for true edge localisation due to noise. The proposed technique offers significant improvement in edge localisation for an image as compared with Canny and Sobel methods.
Keywords :
biomedical MRI; edge detection; mean square error methods; medical image processing; Canny methods; MRI; MSE; Sobel methods; correction factor; edge detection method; edge localisation technique; magnetic resonance imaging; mean square error; noise function; signal dependent Rician noise; signal dependent noise;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
DOI :
10.1049/el.2015.0774
Filename :
7163391
Link To Document :
بازگشت