  • Abstract
    Almost all external radiotherapy planning systems use dose profile data. There are various methods to measure these profiles. Direct measurements with semiconductor detectors are widely used. But this method takes lots of time especially at linear accelerators. We developed a simple method for this purpose. Dose profiles are measured using radiographic films and verification films. Suitable modifications are done with derived sensitometric curves. Results are compared with original profiles. For the 60Co, this method is found as an alternative to other methods with acceptable accuracy (% ±2).
  • Keywords
    dosimetry; radiation therapy; radiography; 60Co; acceptable accuracy; dose profile data; external radiotherapy planning systems; film dosimetry; linear accelerators; radiographic films; sensitometric curve; verification films; Crystals; Density measurement; Detectors; Dosimetry; Imaging phantoms; Optical films; Optical sensors; Radiography; Semiconductor films; Silver;