DocumentCode
1982539
Title
Determination of dose profile data with film dosimetry
Author
Gokce, Tuncay C. ; Aytac, Sitki ; Gokce, Tumay
Author_Institution
Dept. of Comput. Eng., Izmir Inst. of Technol., Turkey
Volume
4
fYear
2001
fDate
2001
Firstpage
3923
Abstract
Almost all external radiotherapy planning systems use dose profile data. There are various methods to measure these profiles. Direct measurements with semiconductor detectors are widely used. But this method takes lots of time especially at linear accelerators. We developed a simple method for this purpose. Dose profiles are measured using radiographic films and verification films. Suitable modifications are done with derived sensitometric curves. Results are compared with original profiles. For the 60Co, this method is found as an alternative to other methods with acceptable accuracy (% ±2).
Keywords
dosimetry; radiation therapy; radiography; 60Co; acceptable accuracy; dose profile data; external radiotherapy planning systems; film dosimetry; linear accelerators; radiographic films; sensitometric curve; verification films; Crystals; Density measurement; Detectors; Dosimetry; Imaging phantoms; Optical films; Optical sensors; Radiography; Semiconductor films; Silver;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE
ISSN
1094-687X
Print_ISBN
0-7803-7211-5
Type
conf
DOI
10.1109/IEMBS.2001.1019699
Filename
1019699
Link To Document