Abstract
The performances of the Michelson (MI) and Mach-Zehnder interferometers (MZI) are compared for laser linewidth measurement with the delayed self-heterodyne technique. Voigt fitting is used to extract the Lorentzian and Gaussian components for the MZI and MI with faraday mirrors and also standard gold mirrors. The MI with Faraday mirrors provides a measurement with significantly lower error, owing to the elimination of noise associated with random polarisation drifts.
Keywords
Mach-Zehnder interferometers; Michelson interferometers; laser mirrors; laser variables measurement; spectral line breadth; Gaussian components; Lorentzian components; Mach-Zehnder interferometer; Michelson interferometer; Voigt fitting; delayed self-heterodyne technique; faraday mirrors; laser linewidth measurements; standard gold mirrors; Distributed feedback devices; Fiber lasers; Laser noise; Measurement by laser beam; Mirrors; Semiconductor lasers;